Determination of proper peaking time for Ultra-LEGe detector
Access
info:eu-repo/semantics/closedAccessDate
2011Access
info:eu-repo/semantics/closedAccessMetadata
Show full item recordCitation
https://doi.org/10.1016/j.radmeas.2011.01.023Abstract
Reducing count losses and pile-up pulse effects in quantitative and qualitative analysis is necessary for accuracy of analysis. Therefore, the proper peaking time for particular detector systems is important. The characteristic X-rays emitted from pure some elements were detected by using an Ultra-LEGe detector connecting a Tennelec TC 244 spectroscopy amplifier at different peaking time modes. Overall pulse widths were determined by an HM 203-7 oscilloscope connecting amplifier. The proper peaking time for Ultra-LEGe is determined as 3.84 mu s. (C) 2011 Elsevier Ltd. All rights reserved.